DSO138 oscilloscope connection guide
DSO138 oscilloscope DSO138 13803K / 13804K wiring guide with power, signal assignments, and bench checks.
DSO138 13803K / 13804K
7 contacts Current packagePower the oscilloscope
RequiredUse a regulated 9 V DC adapter with a center-positive 5.5 × 2.1 mm plug and a current rating above 200 mA.
- Adapter positive
- DC_IN (J10 center)
- Adapter +
- Adapter return
- DC_RETURN (J10 sleeve)
- Adapter −
Bench checks
- Verify the polarity and the unloaded adapter voltage before inserting the plug. The kit accepts 8–12 V DC.
- Power on with the BNC probe disconnected and set coupling to GND. Confirm that the display starts and shows a steady zero trace.
JYE Tech DN138-01v05, Operation and Probe Calibration on p. 3 and schematic on p. 4.
Connection scenarios
Align zero and compensate the analog input
Set coupling to GND and use SEL to highlight VPos. Hold OK for about two seconds to align zero. Then connect the red clip of the BNC probe to J2 and leave the black clip free as shown in the manual.
- Onboard square wave
Bench checks
- Select AC or DC coupling and 0.2 ms/div. Set SEN1 to 0.1 V and SEN2 to ×5. Adjust the trigger level until the waveform is stable, then trim C4 for flat tops and clean corners.
- Set SEN1 to 1 V and SEN2 to ×1, then trim C6 for the same square-wave shape.
JYE Tech DN138-01v05, Operation and Probe Calibration on p. 3 and schematic on p. 4.
DSO138 13803K / 13804K wiring map
J2 · TESTSIG
Onboard 1 kHz, 3.3 V square wave.
Models and documents
This map covers original JYE Tech kits 13803K and 13804K. The optional J4 Mini-USB footprint is listed without a user function. Use J10 or J9 for power.
- DSO138 13803K / 13804K
- JYE Tech kit connectors for the BNC input, DC supply and test signal.
- JYE Tech DSO138 user manual· JYE Tech DN138-01v05, Operation and Probe Calibration on p. 3 and schematic on p. 4.
- JYE Tech DSO138 product documentation· Kit revisions, input characteristics and DC connector dimensions.